| Parameter | Linearity Range | Precision (CV%) |
| WBC(109/ L) | 0.0-99.9 | 2.5(7.0-15.0) |
| RBC(1012/L) | 0.0-9.99 | 2.5(3.50-6.00) |
| HGB(g/L) | 0-300 | 1.5(110-180) |
| MCV(fL) | 40.0-150.0 | 0.5(80.0-110.0) |
| PLT(109/L) | 0-999 | 5.0(200-500) |
| Whole blood | Prediluted | |
| WBC/HGB | 1:333 | 1:676 |
| RBC/PLT | 1: 42143 | 1:85858 |

■ Menu Review, Test, Maintenance, QC, Setup, Shutdown, Prime reagents, System diagnosis, Aperture rinse, Soak and rinse, Calibration Instrument config. ■ Sampler Volume Predilute, 20μL Whole blood, 15μL ■ Aperture Diameter WBC, 80μm RBC/PLT, 70μm ■ Carryover WBC, RBC, HGB<0.5%, PLT< 1%■ Input/Output 1 RS232, 1 keyboard ■ Printout Built-in thermal printer (2 optional print modes) ■ Operating Environment Temperature15℃-30℃ Humidity 30%-85% ■ Power Requirement AC100-240V,50/60Hz ■ Dimension 440mm(H)×350mm(W)×440mm(D) ■ Net Weight 17kg